GE Healthcare · MRSigna22445882226629224487422446032244605
SGA Power Supply Under Internal Wiring Fault or Control Undervoltage Due to SGA Power Supply.
Symptom
Will Not Scan with Gradient Errors.
Error message
2244588 The GP is reporting SGA Power Supply Power Off Fault. Dec 4 Tue 06:21:29 2244588 NSP SCP:Grad 739 The GP is reporting SGA Power Supply Power Off Fault.-> 2226629 Gradient Identification Problem The gradient subsystem could not identify itself as 8915 or 8905 gradients. 2226629 Gradient Identification Problem. The gradient subsystem could not identify itself as 8915 or 8905 gradients. Possible gradient amplifier problem, please run diagnostics. 2244874 The GP reported multiple SGA Power Supply Internal Wiring Faults or Control Undervoltages. the same error as 2244603 on 11X, G3 and 2244605 on 12X, E2, 14M4 (HDx, HDe), and openMR4,5. 2244874 The GP reported multiple SGA Power Supply Internal Wiring Faults or Control Undervoltages. 2244874 error code on Post 14M5 is the same error as 2244603 on 11X, G3 and 2244605 on 12X, E2, 14M4 (HDx, HDe), and openMR4,5. 2244874 error code on Post 14M5 is the same error as 2244603 on 11X, G3 and 2244605 on 12X, E2, 14M4 (HDx, HDe), and openMR4,5.2244874 The GP reported multiple SGA Power Supply Internal Wiring Faults or Control Undervoltages.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 2334862-2 ACGD SGA PS ASM W/FRU PACKAGING See Long Description 2334862-2 Part Type: GP REPAIRABLES Part Weight: 2.18 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ACGD SGA PS ASM W/FRU PACKAGING FRU: Yes Status: Orderable This supply is for use with HFD, HFD-S, and ACGD Cabinets. ** WARNING - POSSIBLE PERSONAL INJURY. HEAVY OBJECT - THE POWER SUPPLY WITH CASE WEIGHS APPROXIMATELY 240 LBS (109 KG). Assistance to lift this product mechanical (Dollies) or Additional personnel] is required. 212002950 May 27, 2009
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.