Pular para o conteúdo principal
GE Healthcare · CTLightSpeed185111

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 185111, HW error: Tilt Interference/Limit switches are INVALID See LD for complete error Mon Nov 28 14:39:35 2005 Error: 185111 Host: ETC Process: Control File: etcValidIntfMatrix.c 1.8 7/9/ Line: 133 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 1066 Log Series: 1 HW error: Tilt Interference/Limit switches are INVALID. Interference/Limit switches = 0xd2db. BCK LIMIT: 1 FWD LIMIT: 0 FWD INTF: 0 BCK INTF: 0

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2268222 TILT POT/SWITCH HARNESS see Long Description for details Part #: 2268222 Part Type: GP CONSUMABLES Part Weight: 0.50 LBR Repairable: No Returnable: No Hazardous Material: N Description: TILT POT/SWITCH HARNESS FRU: Yes Status: Orderable Comments: FLS - TILT POT/SWITCH HARNESS MS2332 05/21/02 NORTECH 4050 NORRIS COURT NW BEMIDJI MN PDR EXISTS
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues