Collapsed Images Due to STIF Board.

Symptom

Scan Pauses with Gradient Errors and Continues after start Scan Pressed then Images Show Bright Line Artifact System scans and pauses with Gradient errors. Images had a definite gradient collapsed.

Error message

2224666 Auto Prescan failed, Auto Prescan failed, possible data acquisition failure. 2224666 mmp-mr01 psc PscMainControl.cc 1311 Auto Prescan failed, possible data acquisition failure.

2248108 APS received data over range error from DRF, on chan= n. 2248108 TOiMR2 NSP APS:dataMoverTask 604 Exception Class: Idle Severity: Unknown Function: Acquisition/Pulse Generation: Acquisition PSD: Spin Echo Coil: US SHOULDER PA Scan: 379/4 APS received data over range error from DRF, on chan = 2 Need to reduce receiver gain.

2244851 The GP is reporting a Gradient Amplifier %c Axis Current Distortion Fault. the same error as 2244577 on 11X, G3, openMR4 and 2244578 on 12X, E2, 14M4 (HDx, HDe), and openMR5. See Long Description 2244851 error code on Post 14M5 and 12.0 M5, is the same error as 2244577 on 11X, G3, openMR4and 2244578 on 12X, E2, 14M4 (HDx, HDe), and openMR5.2244851 The GP is reporting a Gradient Amplifier %c Axis Current Distortion Fault Example error message (could be any gradient axis, Y-axis is shown in this example): Error: 2244851 Host: lx-mr2 Process: NSP File: SCP:Grad Line: 2851 The GP is reporting a Gradient Amplifier Y Axis Current Distortion Fault.-> Extended error messa

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2280952-3 STIF Board See Long Description SRF TRF I/O or STIF 2280952 crosses up in FEMC to 2280952-3. Part Description is "STIF Assembly".
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.