Pular para o conteúdo principal
GE Healthcare · CTLightSpeed18331450013185557183334

LightSpeed 4.X - Octane

Symptom

Scanner Hardware Stopped Scan message DURING Scan data acquisition. Long Description for Feedback Only. Scanner Hardware Stopped Scan message DURING Scan data acquisition.

Error message

Error: 183314 Failure occurred downloading characterization table to the HEMRC. See long description for more details Mon Nov 6 21:07:53 2006 Error: 183314 Host: OBC Process: Generator File: hemrc_manager.c 1.14 Line: 714 Function: XRAY GENERATION : Tube Rotor Control Scan: 50013/1/0 Type: None/Unknown Exception Level: Secondary Time: 21:07:50:815 Log Series: 27 Failure occurred downloading characterization table to the HEMRC.

Error: 185557- hemrc serial link is down. Hemrc_+12V failure detected. Possible causes: HEMRC fuse (F5), Interconnects, HEMRC AC Drive, HEMRC Control Board See Long Description for complete error Mon Jul 24 08:41:32 2006 Error: 185557 Host: OBC Process: Generator File: scanbusDriver.c 1.5 Line: 243 Function: XRAY GENERATION : Tube Rotor Control Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 08:41:28:915 Log Series: 31 Could not send HEMRC message in the allotted time. HEMRC serial link is down. HEMRC_ISO_+12V Failure detected. Possible causes: HEMRC fuse (F5), Interconnects, HEMRC AC Drive,

Error: 183334 HEMRC fault could not be reset See Long Description for complete error Wed Sep 20 10:48:20 2006 Error: 183334 Suite: SMCT Host: OBC 

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.