GE Healthcare · MRSigna2257118
CWA Unable to Download or scan the Propeller PSD.
Symptom
Unable to Download or scan the Propeller PSD Unable to download or scan the Propeller PSD unless the TPS has been reset. Nearly every Propeller scan fails. Sometimes unable to scan other PSDs after a Propeller scan has been run, until the TPS has been reset.
Error message
2257118 Internal Software Error. A reboot is required to recover. See Long Description 2257118 lx-mr scn ScanRMIMgr.cc 4791 Exception Class: Unknown Function: Host: LX MX Interface Internal Software Error. A reboot is required to recover.
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.