GE Healthcare · CTLightSpeed183262
LightSpeed 4.X - GRE
Symptom
Fastcals failed to complete. Fastcals failed to complete
Error message
Error: 183262 Backup contactor opened without firmware command See the long description for the full error Mon May 17 12:20:18 1999 Error: 183262 Suite: CT01 Host: OBC Process: Generator File: ctvrc_manager.c 1.8 Method: No Method Line:823 Function: XRAY GENERATION : 550 VDC Backup Contactor Intlk Scan: 7073/3/15 Type: Axial Exception Level: Pri/Most Severe Time: 12:20:18:495 Log Series:513 Backup contactor opened without firmware command.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2139289-2 CPDU Control Bd. Replaces part 2139289 Used In CPDU - 2269902. 2139289-2 CPDU Control Bd. for Compact PDU Part #: 2139289-2 Part Type: GP REPAIRABLES Part Weight: 5.35 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: 2139289-2 : PDU CONTROL BOARD W/ REW FRU: Yes Status: Orderable Comments: FLS - PDU CONTROL BOARD REWORK ASSEMBLY. THIS P/N 2139289-2, IS P/N T3202KF (2139289) WITH A REWORKED SUB-ASSEMBLY. ADDED PER PCN 220525 BY MS20544 ON 7/17/02. 9/10/02- 2139289-2 REPLACES 2139289 PER PCN 220525. MS12464 11/08/04 SSO# 212001261 FE's are to return defectives and unused/consigned back via FED Ex over night Item #16 A6 panel is the control board
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required