GE Healthcare · CTLightSpeedVCT2100004342005020032100004413863552
LightSpeed 4.X - GRE
Symptom
Scanner Hardware Stopped Scan with Image Allocation errors. Scanner Hardware Stopped Scan with Image Allocation errors
Error message
Error: 210000434 IMAGE ALLOCATION has returned null linkedTuple or null linkedTuple->data See Long Description for complete error Fri Apr 21 07:41:23 2006 Error: 210000434 Host: lightspeed Process: dms File: SaveContext.cxx Line: 184 I MAGE ALLOCATION has returned null linkedTuple or null linkedTuple->data.
ALERT: I/O error in filesystem ("/usr/g/sdc image pool2") meta-data dev 0xed block 0xa00040 ("xfs trans read buf") seen in /var/adm/SYSLOG see LD ALERT: I/O error in filesystem ("/usr/g/sdc image pool2") meta-data dev 0xed block 0xa00040 ("xfs trans read buf") seen in /var/adm/SYSLOG The I/O error could report to a number of different file systems / disk drives. i.e. ("sda8")
Error: 200502003, Cannot allocate image space for XXX images, allocated XX images. See LD for complete error Mon Feb 13 16:36:28 2006 Error: 200502003 Host: ls16dn Process: retroRecon File: srx_sma_manage_image_space.c Line: 690 Function: No system function reported Cannot allocate image space for 148 images, allocated 7 images.
Error: 2052 error during execution of the fonction (note: may or may not be misspelled) See LD for details Tue Jun 7 12:27:18 2005 Error: 2052 hostid: su101 process: imserver imsApplies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2368697-3 HEWLETT-PACKARD HP8000 Computer used in Global Recon Engine (GRE) CONSOLE 2368697-3 GRE CONSOLE HP8000 2368697-3 HP8000 COMPUTER USED IN GRE CONSOLE Part #: 2368697-3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: LINUX DDR WORKST'N PS2 MOUSE--100 PC FRU: Yes Status: Orderable Comments: 3/4/05 (212001261 Shawn Fox) HOST COMPUTER (HP). THISIS THE GOC2 HP 8000 COMPUTER. ***EXTREMELY IMPORTANT: NEW SOFTWARE, S/W LOAD PROCEDURE AND S/W OPTION KEYS ARE REQUIRED IF THE HP8000 COMPUTER IS REPLACED. Please contact the OLC for assistance with new Option Keys
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.