GE Healthcare · CTLightSpeed2687225588
LightSpeed 4.X - GOC2
Symptom
Applications software fails to boot up completely and automatically shuts itself back down. Applications software fails to boot up completely and automatically shuts itself back down
Error message
Error: 1031 stored procedure failed See Long Description for Details. Sun Nov 14 20:00:37 2004 Error: 1031 Host: mcas Process: dbrserver stored procedure failed : -284 - getexamuidinfo failed Also check usr/g/ctuser/logfiles/epdlog with command: tail -100 /usr/g/ctuser/logfiles/epdlog | more MESSAGE from Process 26872>> Sun Nov 14 20:12:15 2004 (netclient.c:93) [Client]> The request manager started MESSAGE from Process 26872>> Sun Nov 14 20:12:48 2004 (:0) [EPD] Failed to remove the exam entry Thu Jul 17 13:45:18 2008 Error: 1031 Host: ct99 Process: dbrserver File: batch.ec Line: 599 stored procedure failed : -243 - Fetch failed for image - ISAM code : -143 Wed Jul 2 10:54:38 2008 Error: 1031 Host: CT01 Process: dbwserver File: batch.ec Line: 934 stored procedure failed : -25588 - Identifyimage failed - ISAM code : 22, object , series $
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GOC2 405L.8_H4.2M4 LightSpeed 4.X with GOC2 and MDAS 16 Gantry H4 Gantry Model # 2339985 and Global Operator Console 2 Model #'s 2366438 and 2366438-2(Hino) with HP8000 CPU (or newer) and Pegasus Recon ICE Box Product: LightSpeed16 405L.8_H4.2M4 CPU: GenuineIntel Intel(R) Xeon(TM) CPU 2.66GHz 4 2657 MHz i686 Processors Main memory size: 2048 Mbytes [root@presbyct2]# sg_map -i /dev/sg0 /dev/sda SEAGATE ST336753LW HPS2 /dev/sg1 /dev/sdb SEAGATE ST373453LW HPS1 /dev/sg2 /dev/sdc SONY SMO-F551-SD 1.05 /dev/sg3 /dev/scd0 MATSHITA DVD-RAM SW-9574S AZX8 /dev/sg4 /dev/scd1 HL-DT-ST DVD-ROM GDR8163B 0B15 [root@presbyct2]# df -h Filesystem Size Used Avail Use% Mounted on /dev/sda2 4.0G 1.7G 2.3G 41% / /dev/sda1 27M 12M 14M 44% /boot none 1008M 0 1007M 0% /dev/shm /dev/sda6 14G 3.5G 10G 25% /usr/g /dev/sda7 9.8G 753M 9.0G 8% /usr/g/sdc_image_pool /dev/sdb1 68G 28G 40G 41% /usr/g/sdc_image_pool2 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 405L.8_H4.2M4 for LightSpeed 4.X (LightSpeed 16) with Console GOC2 (w/Linux PC) (2315653) or Xtream (GRE) Console 405L.8_H4.2M4 for LightSpeed 4.X (LightSpeed 16) with Console GOC2 (w/Linux PC) (2315653) or Xtream (GRE) Console 405L.8_H4.2M4 for LightSpeed 4.X (LightSpeed 16) with Console GOC2 (w/Linux PC) (2315653) or Xtream (GRE) ConsoleClass C Application Software Set CD Part No. Ver. CD Label CD Part No. Ver. CD Label Set Part No. Apps. CD Part No Apps. CD Label showprods/ Service Desktop Intro. Date Comment 2394897) 1104L.3_ H4.2CLASSC LightSpeed/ HiSpeed QXi Class C Software M4 5111370 5111371 LightSpeed 4.X Applications Software 405L.8_H4.2M4 FW11-04 FMI 25338
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.