GE Healthcare · CTLightSpeedVCT230023003200910004
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 230023003 FAILED to resume scanning Scanner Hardware is not ready or unavailable See Long Description for error detail Sat Sep 4 20:52:04 2010 Error: 230023003 Host: swct Process: dailyPrepRx File: DailyPrepUtils.c Line: 4772 FAILED to resume scanning. Possible Reasons: Scanner Hardware is not ready or unavailable. Please verify and try again Error: 200910004 Failed to acquire Scan Hardware Resource. Critical Error. See the Long Description for the full error code Tue Apr 3 11:13:22 2007 Error: 200910004 Host: kghcate_vct Process: dailyPrepRx File: SrvScanActions.c Line: 1091 Failed to acquire Scan Hardware Resource. Critical Error.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5159833 VDARC - Volume Data Acq Recon Control Computer, 2nd Edition (Jarrell). See Long Description for Details.. Part #: 5159833 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: VDARC - Volume Data Acq Recon Control Computer, 2nd Edition FRU: Yes Status: Orderable Parts Offered as: Current : 5159833 Harvested : Not Available Exchange : Not Available Comments: Per Jarrell RSPL, 7/17/06, ms11892. ** FE Assist ** Must order software release 06MW29.7, p/n 5193989 and perform a LFC to support this FRU if system is not currently running 06MW29.7 or better.(12/8/06 212001261
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required