GE Healthcare · MRSigna222482922491102224838
Loose connection on cables at SRI causing a checksum error in the downloaded SRI Code.
Symptom
SRI Communication Error. SRI failed to respond to a packet from the SCP SRI Communication Error. SRI failed to respond to a packet from the SCP.
Error message
2224829 SRI-SCP Communication Error. Wed Apr 28 16:58:01 2004 Error: 2224829 Host: GEMS Process: NSP File: SCP:SRI Line: 200 SRI-SCP Communication Error. A TPS Reset may be necessary. The SRI received a packet from the SCP containing an unknown opcode. 2249110 SRI Communication Error. SRI failed to respond to a packet from the SCP. First packet opcode waiting for a response: 0x47 2224838 SRI Memory Error The SRI detected a checksum error in the downloaded SRI Code
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.