GE Healthcare · MRSigna224860922471272252349224457022448482244574
Site Air Conditioning failure
Symptom
Unable to scan with message System Cabinet Temperature is Greater Than 50'C See long description This message is seen in the message block area, above the six icons on the scan desktop. Soon the message is seen to change to "System Cabinet temperature has returned to 40'C", and the process repeats several times (cycles back and forth between 50'C and 40'C). Reference PQR 13080986
Error message
2248609 The air temperature in the System Cabinet is greater than 50'C (122F) Mon Aug 22 12:32:49 2005 Error: 2248609 Host: MR3T Process: NSP File: SCP:IrfPciHWO Line: 1234 The air temperature in the System Cabinet is greater than 50'C (122F). Check equipment room temp, if warmer than normal contact site maintenance, else call GE Service. diags error is 2247127 2252349 The Driver Module has detected that Fan #2 is malfunctioning. Error 2252349 NSP SCP:DCB 363 Exception Class: Unknown Severity: Unknown Function: RF: RF Bias PSD: Unknown Coil: Unknown Scan: -1/-1 The Driver Module has detected that Fan #2 is malfunctioning.-> 2244570 The gradient amplifier failed to go to Ready when commanded. 2244570 TOiMR2 NSP SCP:Grad 1106 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Spin Echo Coil: US SHOULDER PA Scan: 379/4 The X Axis gradient amplifier failed to go to Ready when commanded.-> 2244570 TOiMR2 NSP SCP:Grad 1106 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Spin Echo Coil: US SHOULDER PA Scan: 379/4 The X Axis gradient amplifier failed to go to Ready when commanded. 2244848 The GP is reporting a Gradient Amplifier %c Axis OverC
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.