Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT200610001290009027

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

ECG/EKG trace / Rpeak editor is not seen in retro recon screen. While doing the acquisition the traces were seen. Please see long description for the detailed issue. During Patient scan the ECG trace was seen. While doing the retro recon the Rpeak editor and ECG trace is not seen

Error message

Error: 200610001 Error message recieved from ECG Viewer. Please see EKGViewer.java.log for more details See Long Description For Details Wed Sep 8 10:23:43 2010 Error: 200610001 Host: ct01 Process: EKGBroker File: EKGBroker.cxx Line: 1307 Error message recieved from ECG Viewer. Please see EKGViewer.java.log for more details

Error: 290009027 Timeout waiting for Rpeaks for ISR_ID: 19 Scan: E21317/S2/S1 See Long Description for Details Wed Sep 8 10:23:44 2010 Error: 290009027 Host: darc Process: recon_control File: ImageSetRequest.cxx Line: 8649 Timeout waiting for Rpeaks for ISR_ID: 19 Scan: E21317/S2/S1 recon_group_scan_index: 1 recon_group_num_scans: 1

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5198566 cable-gantry - EKG PRO See Long Discription for Details Part #: 5198566 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: cable-gantry - exg pro FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.