GE Healthcare · MRSigna
Known Software issue with 15M4B coding error.
Symptom
Install In Spec Does Not Pass with a “EPT GROUP DELAY” Failure see long description KNOWN ISSUES AND WORKAROUND WITH THE HDxt M4B RELEASE · INSTALL IN-SPEC DOES NOT PASS WITH A “EPT GROUP DELAY” FAILURE ISSUE: This issue is ONLY applicable to first time installations on BRM and CRM systems. Install In Spec reports a failure regarding the Group Delay file created by running EPT. This file does not get copied to the correct directory.
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.