GE Healthcare · CTLightSpeed260115206506528101635920390260115196516886
LightSpeed 5.X (80/100kW)
Symptom
Table will not move up or down during a elevation table reference. Table will not move up or down during a elevation table reference
Error message
Error: 260115206 Elevation moved without command or position changed by characterization. See LD for more details Sat Nov 25 07:59:43 2006 Error: 260115206 Host: tgp Process: tgp_subsys File: mtcb_emc_motion_check.c Line: 74 Sat Nov 25 07:59:44 2006 Host : Table Sub-System Ermes # : 260115206 Exception Class : Secondary Severity : Pri/Soft File : mtcb_emc_motion_check.c Line# : 74 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Elevation moved without command or position changed by characterization. Current position = 506528 um Last stop position = 1016359 um Tolerance = 5000 um Time since last stop = 20390 msec Hardware status = 0x7088 Target position = 506528 um Error: 260115196 Elevation lower limit switch is activated. see long description for details Sat Mar 15 06:13:29 2008 Error: 260115196 Host: tgp Process: tgp subsys File: mtcb emc isrfunc.c Line: 79 Sat Mar 15 06:13:29 2008 Host : Table Sub-System Ermes # : 260115196 Exception Class : Secondary Severity : Pri/Soft File : mtcb emc isrfunc.c Line# : 79 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Elevation lower limit switch is activated. Position is 516886 um from
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2342729-3 Elevation or Cradle Drive Amplifier used on the 5.X table - same driver amp is used for both See LD for additional part information Part #: 2342729-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: H-POWER TABLE SERVO DRIVE, EMC EDITION 2 COMPLIANT FRU: Yes Status: Orderable Comments: **Please consider ordering the used part when it is appropriate for your customer. It will save you substantial money.** Part added per DST-HP60 16 slice RSPL. 040906 MS13452.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.