GE Healthcare · CTLightSpeed260003158
LightSpeed 5.X (60kW)
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260003158 The Filter Home Flag Has Moved Since The Last Reference See long description for full error Fri Feb 17 17:56:41 2006 Error: 260003158 Host: HSC Process: CCB Firmware File: filter_home.c Line: 193 Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 8650/2/0 Type: None/Unknown Exception Level: Sec/Soft Time: 17:56:41:390 Log Series: 564 The Filter Home Flag Has Moved Since The Last Reference. Possible Causes: Filter Home Switch, Filter Encoder, Filter Encoder Set Screw. Difference: 360 degrees
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2159578-3 collimator control board used on 1X/2X/3X/4X/5X See Long description for details Part #: 2159578-3 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: COLLIMATOR CONTROL BOARD FRU: Yes Status: Orderable Comments: Not Available
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required