GE Healthcare · CTLightSpeed26000737150359260007372260007428
LightSpeed 4.X - GRE
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260007371 DAS Temperature is HIGH. See LD for more details Mon Aug 28 06:39:05 2006 Error: 260007371 Host: orp Process: orp_subsys File: ctemptst.c Line: 549 Mon Aug 28 06:39:04 2006 Host : Das Ermes # : 260007371 Exception Class : Soft Severity : Pri/Soft File : ctemptst.c Line# : 549 Function : No System Function Reported Scan Type : Static Scan: 50359/1/2 DAS Temperature is HIGH. Highest temperature was seen is on Converter Board 26. Measured: 63.0 degrees C. Acceptable Range: 15.0 to 62.0 degrees C. See Following Error Message for Converter Card List Error: 260007372 Data Acquisition System Is Over Temperature. See LD for more details Mon Aug 28 06:39:43 2006 Error: 260007372 Host: orp Process: orp_subsys File: ctemptst.c Line: 534 Mon Aug 28 06:39:41 2006 Host : Das Ermes # : 260007372 Exception Class : Pause Severity : Secondary File : ctemptst.c Line# : 534 Function : No System Function Reported Scan Type : Static Scan: 50359/2/1 Data Acquisition System Is Over Temperature. Error: 260007428 WARNING: DAS Temperature is nearing the Upper Limit. See LD for more details Mon Aug 28 06:39:06 2006 Error: 260007428 Host: orp Process: orp_subsys File: ctemptst.c Line: 600 Mo
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5324279 U0522AM_FAN AXIAL,172x150x51 5324279 U0522AM_FAN AXIAL,172x150x51 Part #: 5324279 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: U0522AM_FAN AXIAL,172x150x51 AC115 FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required