Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260007371 DAS Temperature is HIGH. See LD for more details Mon Aug 28 06:39:05 2006 Error: 260007371 Host: orp Process: orp_subsys File: ctemptst.c Line: 549 Mon Aug 28 06:39:04 2006 Host : Das Ermes # : 260007371 Exception Class : Soft Severity : Pri/Soft File : ctemptst.c Line# : 549 Function : No System Function Reported Scan Type : Static Scan: 50359/1/2 DAS Temperature is HIGH. Highest temperature was seen is on Converter Board 26. Measured: 63.0 degrees C. Acceptable Range: 15.0 to 62.0 degrees C. See Following Error Message for Converter Card List

Error: 260007372 Data Acquisition System Is Over Temperature. See LD for more details Mon Aug 28 06:39:43 2006 Error: 260007372 Host: orp Process: orp_subsys File: ctemptst.c Line: 534 Mon Aug 28 06:39:41 2006 Host : Das Ermes # : 260007372 Exception Class : Pause Severity : Secondary File : ctemptst.c Line# : 534 Function : No System Function Reported Scan Type : Static Scan: 50359/2/1 Data Acquisition System Is Over Temperature.

Error: 260007428 WARNING: DAS Temperature is nearing the Upper Limit. See LD for more details Mon Aug 28 06:39:06 2006 Error: 260007428 Host: orp Process: orp_subsys File: ctemptst.c Line: 600 Mo

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5324279 U0522AM_FAN AXIAL,172x150x51 5324279 U0522AM_FAN AXIAL,172x150x51 Part #: 5324279 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: U0522AM_FAN AXIAL,172x150x51 AC115 FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution