Pular para o conteúdo principal
GE Healthcare · MRSigna2249418222385222238802249436

Cannot boot applications or reset TPS with Errors 2223852 2223880 2249418 2249436 due to ICN.

Symptom

Application Software does not complete the boot process with TPS Download failure

Error message

2249418 Network problem detected. Check MGD subnet network configuration. See Long Description Wed Apr 14 09:27:10 2004 2249418 2 CARD diagCoord mgdArp.c 157 Task ID: 0xf99df8 Time: 981 Object: SCP Network problem detected. Check MGD subnet network configuration. etherAddrResolve failed for network interface motfec0, IP 10.47.1.2, numTries 5, delay 100 ticks. status = -1.

2223852 The APS board reported a communication manager (CM) initialization error. See Long Description Please check the error log for a more detailed error message from the board. Extended Error: MCP820 boards can be identified by the word MCP820 printed on the board extraction lever

2223880 Reset TPS was not successful. Board Failure Status APS Fail, SCP: Pass, AGP Pass, VRE Fail. ErrorNo: 9. See Long Description Extended Error: The MGD failed to start a TPS Reset. Either the serial connection to the MGD STIF Board is not working or the Ethernet connection from the Host to the MGD Hub is not working. Run the APS, AGP, and SCP Host Data path ping tests to check the Ethernet connections. Run the APS, AGP, and SCP Board Level Diagnostics to check the SCSI Serial Expansion. The STIF Board serial connection is not t

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
SecondOpinion Image Compute Node 5127452 See comments in Resolution Concept for second opinion requirements before ordering this part.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.