Pular para o conteúdo principal
GE Healthcare · CTLightSpeed290009030200352906290009015200352908200001821

LightSpeed 5.X (60kW)

Symptom

Message popped-up on screen "Reconstruction process has stopped, scanning is possible but no images will be made"

Error message

Error: 290009030 Failed to stop image_generation process See long description for full error Mon Nov 19 06:31:41 2007 Error: 290009030 Host: darc Process: recon_control File: IGControl.cpp Line: 591 Failed to stop image_generation process.

DARC /var/log/messages : timeout attempting to rsh to ig2 ( can be ig1 or ig3 ) See long description DARC /var/log/messages Aug 18 18:43:09 darc logger: is_IG_up 1 ig2 called Aug 18 18:43:09 darc is_IG_up:[recon_control:7973]: ig3 is up ( this will be typically 5 mins after start of DARC boot ) Aug 18 18:43:09 darc is_IG_up:[recon_control:7967]: ig1 is up Aug 18 18:43:10 darc is_IG_up:[recon_control:7970]: timeout attempting to rsh to ig2 ( this is ig2 failing to respond to DARC ) With no intervention DARC will retry to contact ig that failed to respond on fisrt attempt ( notice timestamps ) Aug 18 18:49:58 darc logger: is_IG_up 3 ig2 called Aug 18 18:49:58 darc mountd[3273]: authenticated mount request from ig2:855 for /usr/g (/usr/g) Aug 18 18:49:58 darc is_IG_up:[recon_control:7970]: ig2 is up ( ig2 up nearly 7 minutes after second start )

Error: 200352906 Errors starting up the recon sub system. See Long Description for complete error Mon O

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.