GE Healthcare · MRSigna22246662254782

Unable to Scan with Power Monitor Faults due to Split Head Coil Halves Were Not Completely Closed Together.

Symptom

Unable to scan with Power Monitor Faults

Error message

2224666 Auto Prescan Failed, RF Amplifier is Not Ready NOTE: When using this error to create new knowledge, please include any associated error codes. 2224666 is a generic error code which requires further information. Mon May 17 08:19:08 2004 Error: 2224666 Host: lx_mr Process: psc File: PscMainControl.cc Line: 1296 Auto Prescan failed, RF amplifier is not ready.

2254782 The Universal Power Monitor 1 has detected Peak Power Exceeded on the Narrow Band Body Amplifier. Sun Nov 20 12:49:19 2005 Error: 2254782 Host: HDMR Process: NSP File: SCP:UPM Line: 657 The Universal Power Monitor 1 has detected Peak Power Exceeded on the Narrow Band Body Amplifier. Peak value detected = 2236

2254782 The Universal Power Monitor 2 has detected Peak Power Exceeded on the Narrow Band Head Amplifier. 2254782 1 df15hdmr NSP SCP:UPM 657 Exception Class: Unknown Severity: Unknown Function: RF: RF Power Monitor PSD: Spin Echo Coil: HEAD Scan: 7975/1 The Universal Power Monitor 2 has detected Peak Power Exceeded on the Narrow Ban d Head Amplifier. Check patient weight entry. Peak value detected = 657 W. Max Peak Limit = 626 W-

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required