Site Air Conditioning failure.
Symptom
The equipment room is very hot. Customer has called for AC repair, the computer room temp is 93 F. The control room and scan room temps appear to be good.
Error message
2248609 The air temperature in the System Cabinet is greater than 50'C (122F) Mon Aug 22 12:32:49 2005 Error: 2248609 Host: MR3T Process: NSP File: SCP:IrfPciHWO Line: 1234 The air temperature in the System Cabinet is greater than 50'C (122F). Check equipment room temp, if warmer than normal contact site maintenance, else call GE Service. diags error is 2247127
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR