GE Healthcare · X-ray R&FDefinium1090419

OTS drops vertically when using Auto Position

Symptom

OTS drops vertically when using Auto Position See long description. When driving OTS using Auto Positioning, OTS will begin to drive down prior to reaching destination. This usually occurs when driving to Table, but can occur when driving to WS. This tends to occur when driving the OTS using Auto Position from either Park or WS to Table. The OTS will drive up to the calibrated position as normal and will begin to move toward the selected destination. Once the OTS gets close to the final destination, the OTS appears to drop downward. This movement could strike the patient. If this occurs, create and submit an safety report.

OTS drops vertically dirver error

Error message

1090419 - OTS Vertical Amp is not alive

Applies to

Models
Definium
Product
Definium 8000
Software
All Versions
Component
Wallstand (Thunder)
Area
X-ray R&F

Full cause & step-by-step resolution