GE Healthcare · CTLightSpeed290009018
LightSpeed 4.X - GRE
Symptom
A Pink Pop up message is reported on boot-up that Recon times may be slower - scanning may continue See Long description to see the Pink Pop up message A Pink Pop up message is reported on boot-up that Recon times may be slower
Error message
Error: 290009018 All the IG nodes are not operational. See long description for full error Mon Nov 29 06:50:22 2004 Error: 290009018 Host: darc Process: recon control File: recon control cb.cxx Line:1549 All the IG nodes are not operational
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5127222-2 Cable, Ethernet, 2ft Lg - CAT5e used in consoles See long description for details Part #: 5127222-2 Part Type: GP CONSUMABLES Part Weight: 2.43 LBR Repairable: No Returnable: No Hazardous Material: N Description: Cable, Ethernet, 2ft Lg - CAT5e, 4 twisted pair, 24awg solid copper, Shielded - aluminized polyester and tinned copper, gray PVC jacket, snagless PVC boot, Plugs - UL94V-0 rated polycarbonate with metallic strain relief, EIA-TIA568, UL list FRU: Yes Status: Orderable Comments: Added per VCT M4 RSPL update, ms13452
- Area
- CT
Full cause & step-by-step resolution