Pular para o conteúdo principal
GE Healthcare · MRSigna2244655

Will Not Scan with Gradient Errors Due to the GSC Twin Gradient Switch Module Box

Symptom

Will Not Scan with Gradient Errors.

Error message

2244655 Gradient Switch Fault Detected. Scan will be paused 2244655 Host: TwinExciteLXMR Process: NSP File: SCP:Grad Line: 704 Gradient Switch Fault Detected. Scan will be paused.-> Extended Error: Please restart the scan (continuing the scan might cause image quality issues). If the fault persists,replace the Gradient Switch

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2253466 TWIN GRADIENT SWITCH Module Box Same part number for 1.5T and 3.0T. Same part number for 1.5T and 3.0T.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues