GE Healthcare · MRSigna222593922253372225704
No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet No Power to the Tabl…
PDU was in stop mode
Symptom
No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
Error message
2225939 The PAC did not respond correctly to a reset command from IPG. Error: 2225939 Host: rmr2 Process: NSP File: HPC_PORT_ID:spu_os_init Line: 469 The PAC did not respond correctly to a reset command from IPG. Check PAC power and fiberoptic connections, run PAC diags. 2225337 SPI-SRI Communication Error. SRI failed to respond to a watchdog packet from SPI. Error: 2225337 Host: MRS2 OC0 Process: NSP File: HPC PORT ID:spi sri watchdog fail Line: 13 SPI-SRI Communication Error. SRI failed to respond to a watchdog packet from SPI. Watchdog packets waiting for a response: 0x0001 2225704 MDS link failure. The MDS fiber optic link appears to be broken between the RF and System cabinets. Error: 2225704 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_mds Line: 300 MDS link failure. The MDS fiber optic link appears to be broken between the RF and System cabinets. Check RF cabinet power, fiber optic connections and signal level of the fiber optic link between the cabinets. Run MDS fiber optic link diagnostics.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.