Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT18415150004

LightSpeed 3.X (Ultra, Plus, QX/i) - GRE

Symptom

"Scan hardware Stopped Scan" with scan data errors. "Scan hardware Stopped Scan" with scan data errors

Error message

Error: 184151 Scan Control timed out waiting for the Ready For Views message from the SRU See the long description for the full error code Sat Apr 17 20:29:44 2004 Error: 184151 Host: STC Process: Scan Control File: scTrigSeqActions.c 1.65 Line: 1365 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 50004/3/1 Type: Axial Exception Level: Pri/Most Severe Time: 20:29:43:355 Log Series: 676 Scan Control timed out waiting for the Ready For Views message from the SRU. This message is sent from the SRU once offset collection has completed. Possible causes: trigger failure, DIP failed to receive data, message lost, software error

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5147442 Jarrell DARC2, used ONLY on GOC4 with 16 SLICE MDAS or 4/8/16 slice GDAS - CT or PET 07mw and above software See Long Description for details. 5147442 Jarrell DARC2 , replaces 2362870-10 DARC - used ONLY on GOC3 or GOC4 with 16 SLICE MDAS or 4/8/16 slice GDAS -emc2 compliant Click here to see all available DARC part numbers Note: 5147442 replacement applies ONLY to CT and PET 07mw and above software, or CT 06mw29.7 Part #: 5147442 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Computer - Data Acquisition Recon Control, 2nd Generation FRU: Yes Status: Orderable Comments: Per Jarrell RSPL, 7/19/06, ms11892. ** FE Assist ** Must order software release 07MW11.10, p/n 5220443 (deployed by FMI 25389, p/n 5191771), and perform a LFC to support this FRU if system is not currently running sw rls 06MW29.7 (6/28/07 212001261) Part #: 2362870 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: DATA ACQUISITION & RECON CONTROL-GRE FRU: Yes Status: Orderable Comments: FLS - ADDED PER GRE RSPL 5.12.03 MS27898 DATA ACQUISITION & RECON USED ON H16 CONTROL-GRE MS13452 030519 RISK: SITE DOWN FREQEUNCY: LOW ROUTED TO RDC FOR REWORK PER UPCOMING PCN. MS12575 2/20/04 PDR REQEUSTED IM05 REQUESTED ## #** FE Assist ** applies to CT Not Pet systems. If no stock is available order a DARC II + ICOM replacement kit per the following * 5144058-500, 16 MDAS 16,8,4 GDAS DARC Replacement kit, JARRELL Direction 5194645-100 , you MUST also order p/n 5193989 software release 06MW29.7 or LATER, and perform a LFC. See Service note SN06101010A.(2/12/07 212001261)### *** 2ND OPINION NOTES *** Reload darc OS & APPS. For reasons unknown darc SW gets corrupt. RSE - If part fails due to a noisy fan please document this with a PQR GOC3 2362870-2 DARC for 4/8 slice MDAS Part #: 2362870-10 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Computer - Data Acquisition Recon Control, 16-slice MDAS and GDAS, 2U, EMC Ed2 Compliant FRU: Yes Status: Orderable Comments: added per EMC2 eco 2007993 as replacement for 2362870, ms13452 041015 ** FE Assist ** applies to CT Not Pet systems. If no stock is available order a Kit 5144058-500 : 16 MDAS 16,8,4 GDAS DARC REPLACEMENT KIT, JARRELL Refer to Dir 5194645-100, you MUST also order p/n 5193989 software release 06MW29.7 or LATER, and perform a LFC. See Service note SN06101010A.(2/12/07 212001261) *** 2ND OPINION NOTES *** DARC failures in system that are caused by noisy p/s fans should have a PQR written to part. IP network address should not be changed unless hospital network is on the same IP subnet. Make sure the proper DARC is ordered to match DAS/Slipring. DOA DARC should be opened and inspected for loose connections. added per EMC2 eco 2007993 as replacement for 2362870 ms13452 041015
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.