Pular para o conteúdo principal
GE Healthcare · CTLightSpeedInnova260125034

Pop up message saying: Warning, Press Fast and Slow buttons 3 seconds to reset.

Symptom

Pop up message saying: Warning, Press Fast and Slow buttons 3 seconds to reset. See long description . This refers to the Angio Control Panel on the ATOM system.

Error message

Error: 260125034 Did not reboot TCB within 20sec. See Long Description for complete error Tue Aug 8 21:50:05 2006 Error: 260125034 Host: tgp Process: tgp_subsys File: ta_func_gt.c Line: 735 Tue Aug 8 21:50:04 2006 Host : Table Gantry Processor Ermes # : 260125034 Exception Class : Secondary Severity : Pri/Soft File : ta_func_gt.c Line# : 735 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Did not reboot TCB within 20sec.

Applies to

Models
LightSpeed, Innova
Product
ATOM CT - ( Advanced Technology On Mobile Gantry ) (ACT) 5.x HP60 LightSpeed 16 CT with Innova 4100 Advantix LCA Plus / LCV Plus Vascular System See Long Description for details ATOM CT - ( Advanced Technology On Mobile Gantry ) (ACT) 5.x HP60 LightSpeed 16 CT with Innova 4100 Advantix LCA Plus / LCV Plus Vascular System What’s the ATOM Project? Clinical Efficacy Application Hepatocellular Carcinoma Hepatic Cancer for interventional procedure as TAE Neuroradiology Neuro intervention Vertebroplasty Pre-intervention Efficiency Decrease the operation time for intervention Prevention of infection inside the House less Co-medical Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required