GE Healthcare · X-ray R&FDefinium109036910002431090453109037010904511090382
System displays a vertical homing error or “AutoPosition is not ready yet” message.
Symptom
Tube dosen't locate at 0 degree after booting
System displays a vertical homing error or “AutoPosition is not ready yet” message. See Long Description System displays a vertical homing error or message that “AutoPosition is not ready yet.” In the message log there will also be a message “The OTS is not responding. Retry or reset the system.”
autopositionning doesn't work after boot
Error message
1090369-OTS Tube Angulation Amplifier Current output limit Error 1000243 - AutoPosition is not ready yet 1090453 - An OTS CAN Command was not processed successfully 1090370 - OTS Tube Angulation Amplifier Voltage output limit Error 1090451 - OTS Homing Failure 1090382 - OTS Tube Angulation Amplifier Velocity limit Error 1090378 - OTS Tube Angulation Amplifier tracking warning
Applies to
- Models
- Definium
- Product
- Definium 8000
- Component
- Definium 8000 OTS
- Area
- X-ray R&F
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray R&F issues
- AEC Backup Time message
- Definium 8000 Auto positioning does not work after a reset/reboot due to wall stand receptor at a hard limit during rebooting or repowering.
- Definium 8000 One or both monitors display the following message on every boot
- Definium 8000 Wallstand or Table Only Configurations running VolumeRAD M3 Software.
- Definium 8000 OTS Collimator has no power due to a tripped breaker (CB8) ont he Tube Suspension Logic (TSL) Board.
- Definium 8000 The Siemens Collimator has not power due to a tripped breaker (CB9) on the Tube Suspension Logic (TSL) Board.