GE Healthcare · CTLightSpeedVCT23000089750074230000065
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Scanner Hardware Stopped Scan With A Pink Pop Up Message " Data Acquisition is not ready for scanning.You might have to reboot the system to recover." See Long Description for Details Scanner Hardware Stopped Scan With A Pink Pop Up Message Is Reported Customer cannot perform scanning, the system report scanner hardware stop scan ,and a pink pop up message is reported that:" Data Acquisition is not ready for scanning.You might have to reboot the system to recover."
Error message
Error: 230000897 Write to scan data disk failed. See LD for more details Tue May 23 07:22:28 2006 Error: 230000897 Host: darc Process: dataacq File: raw_file.cxx Line: 250 Function: Data Acquisition : Scan Data Save Scan Type: Not defined Scan: 50074/1/1 Scan Seq Id: 0 Write to scan data disk failed. Status: -1 Most Recent System Error: OS error : 98 --- Address already in use Error: 230000065 Buffer overrun detected during scan data transfer. See long description for full error Error: 230000065 Buffer overrun detected during scan data transfer. This is generally not a hardware issue and will not be remedied by replacing or otherwise servicing the gantry, fiber-optic cable or DIP card. Interrupt bit: 6 Violation count: 1 Scan aborted.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required