GE Healthcare · CTLightSpeedVCT2600074285035926000742950267260007774
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE
Symptom
Cannot scan with Data Acquisition System Temperature is nearing Upper Limit
Error message
Error: 260007428 WARNING: DAS Temperature is nearing the Upper Limit. See LD for more details Mon Aug 28 06:39:06 2006 Error: 260007428 Host: orp Process: orp_subsys File: ctemptst.c Line: 600 Mon Aug 28 06:39:04 2006 Host : Das Ermes # : 260007428 Exception Class : Soft Severity : Sec/Soft File : ctemptst.c Line# : 600 Function : No System Function Reported Scan Type : None/Unknown Scan: 50359/1/0 WARNING: DAS Temperature is nearing the Upper Limit. Highest temperature was seen is on Converter Board 23. Measured: 63.0 degrees C. Acceptable Range: 15.0 to 62.0 degrees C. See Following Error Message for Converter Card List Error: 260007429 WARNING: Data Acquisition System Temperature is nearing the Upper Limit. See long description for a full sample error message. Wed Jul 16 06:39:30 2008 Error: 260007429 Host: DAS Process: DCB Firmware File: ctemptst.c Line: 584 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 50267/8/56 Type: Axial Exception Level: Pri/Soft Time: 06:39:30:680 Log Series: 779 WARNING: Data Acquisition System Temperature is nearing the Upper Limit. Error: 260007774 List of Converter Cards From the Previous Error Message Card List: xx xx See Long Descriptio
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5324279 U0522AM_FAN AXIAL,172x150x51 5324279 U0522AM_FAN AXIAL,172x150x51 Part #: 5324279 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: U0522AM_FAN AXIAL,172x150x51 AC115 FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required