GE Healthcare · CTLightSpeed2450476010002300190021144566814877214
LightSpeed 2.X (Plus, QX/i) - Octane
Symptom
"Scan hardware Stopped Scan" with scan data errors. "Scan hardware Stopped Scan" with scan data errors
Error message
Error: 245047 Unable to allocate scan files See the long description for the full error code Thu May 12 17:43:37 2005 Error: 245047 Host: rrmcct16 Process: scanmgr File: scan_sequence.cxx Line: 1461 Function: System Control : Scan Control Scan Type: Scout Scan: 0/0/0 Scan Seq Id: 601000 Unable to allocate scan files Error: 230019002 The scan disk I/O layer failed to consume a response from the scan disk server. See LD for complete error Tue Nov 2 03:19:30 2004 Error: 230019002 Host: sgredct1 Process:scanfilemgr File: sdio proxy.c Line:174 The scan disk I/O layer failed to consume a response from the scan disk server. function:consume sequence:6708 raw_data - lost chains " cross-linked with another chain ( found in the iceConsole.log ) See LD for Error Details To View Scan Data Disk Errors in the iceConsole.log cd /usr/g/service/log tail -100 iceConsole.log | more (tail –500 would show the last 500 lines of this log) (use the space bar to page through this log) Look for any Scan Disk Errors as seen below: "/raw_data - lost chains " cross-linked with another chain "/raw_data - lost chains " cross-linked with another chain "/raw_data/ex50752.CT01.1144566814.877214/ov.s7a13" has ille
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2225004-10 73 GB destroked to 18GB 15000 Used on Octane OC with 9Gb OS hard drive. See Long Description for more detail. This part is used on Octane Operators Consoles that have a 9Gb OS hard drive and need a destroked Scan Data Disk. See SN0711061b for jumper settings. Part #: 2225004-10 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: 73 GB destroked to 18GB 15000 RPM Scan Data Drive FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.