Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT260133171260133172

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

"Scan hardware Stopped Scan" with collimator errors. "Scan hardware Stopped Scan" with collimator errors

Error message

Error : 260133171 An error occured attempting to home CAM A or B. See Long Description for Details Wed Jul 29 08:19:20 2009 Error:260133171 Host: orp Process: orp subsysFile: aperture home.cxx Line: 353 Wed Jul 29 08:19:20 2009Host : Collimator Ermes # : 260133171 Exception Class : Abort Severity : PrimaryFile : aperture home.cxx Line# : 353 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 An error occured attempting to home CAM B.

Error: 260133172 The Aperture failed to move to the requested position See Long Description for complete error Thu Feb 23 08:15:31 2006 Error: 260133172 orp orp_subsys aperture_move.cxx 231 Thu Feb 23 08:15:31 2006 Host : Collimator Ermes # : 260133172 Exception Class : Abort Severity : Secondary File : aperture_move.cxx Line# : 231 Function : No System Function Reported Scan Type : None/Unknown Scan: 1352/1/0 The Aperture failed to move to the requested position. EN 399

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5129998-2 ccb chassis assembly VCT-GAN-COLL This is actually the Collimator Control Board and the metal bracket only - it is NOT the complete collimator assembly. See LD for details 5129998-2 ccb chassis assembly VCT-GAN-COLL This is actually the Collimator Control Board and the metal bracket only - it is NOT the complete collimator assembly Part #: 5129998-2 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: CCB Chassis Asm, VCT Failsafe VCT-GAN-COLL FRU: Yes Status: Orderable Comments: Added by ECR 2019627, Part of Failsafe Collimator for VCT, ms13452 051015 This part is direct replacement for 5129998, and 5123185. ms13452
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues