Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT260111604

LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry.

Symptom

Step and Shoot (SmartStep or SmartView) hand controller not functioning See LD for more Step and Shoot (SmartStep or SmartView) hand controller not functioning When customer attempts to use the hand controller for VCT step and Shoot, the system will not initialize into step and shoot mode.

Error message

Error: 260111604 Tx retries used. See long description for a full sample error. Wed Jul 2 11:44:36 2008 Error: 260111604 Host: tgp Process: tgp subsys File: hc tx.c Line: 143 Wed Jul 2 11:44:36 2008 Host : Table Gantry Processor Ermes # : 260111604 Exception Class : Abort Severity : Primary File : hc tx.c Line# : 143 Function : No System Function Reported Scan Type : Axial Scan: 4239/3/1 Tx retries used.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2199945-2 CT SONATA FOOTSWITCH FOR 2X/3X/4X See long Description for details Part #: 2199945-2 Part Type: GP CONSUMABLES Part Weight: 1.24 LBR Repairable: No Returnable: No Hazardous Material: N Description: CT SONATA FOOTSWITCH FOR H2 FRU: Yes Status: Orderable Comments: FLS - CT SONATA FOOTSWITCHFOR H2. THE CABLE LENGTH IS LONGER THAN 2199945. THE -2 P/N IS BACKWARD COMPATIBLE, BUT IT ISBEING CREATED FOR A NEW PRODUCT OFFERING, PER PCN 220511. MS20544 ON 5/24/02. 2199947 CT SONATA HAND HELD CONTROLLER CT SONATA HAND HELD CONTROLLER Part number 2199947 CT SONATA HAND HELD CONTROLLER Part number 2199947
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required