GE Healthcare · CTLightSpeedVCT260107413
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
"Scan hardware Stopped Scan" with Axial Errors. "Scan hardware Stopped Scan" with Axial Errors
Error message
Error: 260107413 The Axial AC Drive reported fault code: F:13 Ground Fault See long description for full error Thu Jul 24 09:08:06 2008 Error: 260107413 Host: tgp Process: tgp_subsys File: AX2fault.c Line: 195 Thu Jul 24 09:08:06 2008 Host : Table Gantry Processor Ermes # : 260107413 Exception Class : Abort Severity : Primary File : AX2fault.c Line# : 195 Function : AXIAL CONTROL : Axial Communications Scan Type : Scout Scan: 193/1/1 %s The Axial AC Drive reported fault code: F:13 Ground Fault Check the motor and external wiring to the drive output terminals for a grounded condition.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 2235342-3 AXIAL MOTOR W/ SPROCKET, HPOWER 2235342-3 AXIAL MOTOR W/ SPROCKET, HPOWER Part #: 2235342-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: AXIAL MOTOR W/ SPROCKET, HPOWER FRU: Yes Status: Orderable The axial motor FRU is item 1 in picture below:
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.