Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT244634244520

LightSpeed 5.X (60kW)

Symptom

Attention: The Scanning Hardware is not available. Please see long description. Attention: The Scanning Hardware is not available, please wait for the hardware reset successfull status posted in the feature status area on the right monitor. If the hardware reset fails, then reset the scan hardware from the Service Desktop System Resets button.

Error message

Error: 244634 Subsystem State; - DCB:down - CCB:up, TCB:up, JEDI:up, TGP/ORP:up, DEVICES:up GOC6 version DCB:DOWN See Long Description Wed Feb 17 23:30:55 2010 Error: 244634 Host: bay46 Process: fwmgr File: fwmgr_utils.c Line: 677 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Subsystem State; DCB:down CCB:up TCB:up JEDI:up TGP/ORP:up DEVICES:up

Error: 244520 Scanning hardware has been shut down. NOTE: This is a very generic error. You need to search the log for a more specific error. Fri Apr 21 07:30:53 2006 Error: 244520 Host: oc01 Process: fwmgr File: HwManager.c Line: 1536 Scanning hardware has been shut down. NOTE: This is a very generic error. You need to search the log for a more specific error.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues