Symptom
Table elevation numbers fluctuate up and down on their own or sometimes may not change at all during elevation Table elevation numbers fluctuate up and down on their own or sometimes may not change at all during elevation.
Error message
Error: 260134707 Elevation could not stop with stop command. If table is moving downward, check valve See Long Description for error versions Mon Sep 19 08:02:03 2005 Error: 260134707 Host: tgp Process: tgp_subsys File: gtcb_emc_movechk.cxx Line: 616 Mon Sep 19 08:02:03 2005 Host : Table Sub-System Ermes # : 260134707 Exception Class : Secondary Severity : Pri/Soft File : gtcb_emc_movechk.cxx Line# : 616 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Elevation could not stop with stop command. If table is moving downward, check valve. last position = 793822 um current position = 794005 um commanded direction = UP time from stop = 587960 ms Mon Dec 12 10:22:24 2005 260134707 Host: tgp tgp_subsys gtcb_emc_movechk.cxx 579 Mon Dec 12 10:22:24 2005 Host : Table Sub-System Ermes # : 260134707 Exception Class : Secondary Severity : Pri/Soft File : gtcb_emc_movechk.cxx Line# : 579 Function : No System Function Reported Scan Type : None/Unknown Scan: 50189/1/0 Elevation could not stop with stop command. If table is moving downward, check valve. last position = 418827 um current position = 418382 um commanded direction = DOWN time from stop = 1040 ms
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5245271 GTCB3 (Global Table Control Board 3) This board replaces 5142177-3 GTCB2 and GTCB part # 5112470 thru 5112470-5, and 5142177. See long description for details Part #: 5245271 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: GTCB3 BOARD ASSY FRU: Yes Status: Orderable Comments: added per to ECO2080741 by sso#305010201 in May 8, 2009. Part will be available end of Sep, 2009, please order old p/n before new p/n on shelf.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.