LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
A pink Attention box pops up during Fast Cal saying there is a problem with the DAS/Detector. Scanning will NOT be disabled. Image Quality will be compromised, call GE SERVICE. See long description for pop up message picture. A pink Attention box pops up during Fast Cal saying there is a problem with the DAS/Detector. Scanning will NOT be disabled. Image Quality will be compromised, call GE SERVICE. After clicking OK Fast Cal completes. ATTENTION: The system detected a problem with the DAS/Detector sub-system. Image Quality will be compromised. Scanning will NOT be disabled, but it is critical that the user call GE SERVICE to rectify the problem. Press 'OK" to proceed at the risk of Image Artifacts
Error message
Error: 230007007 Exclusion Test - Row 33:1B, Channel 899, Module 57 Short together* FAILED PROXIMITY CRITERIA *. Sat Sep 17 08:08:02 2005 Error: 230007007 Host: amhvct Process: badchanmgr File: BadChannelMap.cxx Line: 627 Exclusion Test - Row 33:1B, Channel 899, Module 57 Short together* FAILED PR OXIMITY CRITERIA *. Error: 230007005 Exclusion Test - Row XX:YYZ; Channel XXX; Module YY Short together* FAILED Z-AXIS EXCLUSION CRITERIA *. See Long Description for error example Sat Sep 17 08:08:02 2005 Error: 230007005 Host: amhvct Process: badchanmgr File: BadChannelMap.cxx Line: 615 Exclusion Test - Row 64:32B; Channel 769; Module 49 Short together* FAILED Z-AXIS EXCLUSION CRITERIA *. Error: 230007004 Channel ###, Module ## Short to ground* FAILED X- AXIS EXCLUSION CRITERIA EXCLUSION CRITERIA *. Error: 230007004 can be used for any row, channel and module that is failing. Error: 230007004 Channel ###, Module ## Short to ground* FAILED X- AXIS EXCLUSION CRITERIA Examples: Sat Sep 17 08:08:02 2005 Error: 230007004 Host: amhvct Process: badchanmgr File: BadChannelMap.cxx Line: 584 Exclusion Test - Row 33:1B, Channel 900, Module 57 Short to ground* FAILED X-AXIS EXCLUSION CRITERIA *. T
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5143018 VCT HALO DIFB (Digital Interface Board) See long description for details If no stock of 5143018 Can use P/N 5194006-3 on Halo Detector Part #: 5143018 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: HALO DIFB PWA FRU: Yes Status: Orderable Comments: ** FE ASSIST ** If no stock of 5143018 Can use P/N 5194006-3 on Halo Detector (11/3/09 212001261) Activated per Expert Case/FEMC Case #6717276 212008958 10-25-07.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.