Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26000314950267157995064626000102350723

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

Fastcals failed to complete. Fastcals failed to complete

Error message

Error: 260003149 Illegal distance for CAM A encoder count conversion. Could be Cam B also See LD for complete error codes Fri Oct 12 12:43:36 2001 Error: 260003149 Host: HSC Process: CCB Firmware File: cfg_cam.c Line: 191 Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 50267/1/2 Type: Static Exception Level: Sec/Soft Time: 12:43:36:650 Log Series: 57 Illegal distance for CAM A encoder count conversion. Requested distance: 15799 micrometers Mon Oct 2 10:06:25 2006 Error: 260003149 Host: HSC Process: CCB Firmware File: cfg_cam.c Line: 191 Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 50646/4/40 Type: Axial Exception Level: Sec/Soft Time: 10:06:25:150 Log Series: 969 Illegal distance for CAM B encoder count conversion. Requested distance: -259 micrometers

Do Not Use - use one of the other two Error: 260001023 See Long Description for complete error NOTE: The error message code included in this error is variable. You will need to refer to the log to find the cause of this error. Tue Dec 7 07:23:22 2004 Error: 260001023 Host: OBC Process: HOP File: hop_erlog.c Line: 367 Function: No System Function Reported Scan: 50723/1/1 Type: Axial Exception Lev

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues