Procedure to Adjust the TCU is Incorrect.

Symptom

Procedure to Adjust the TCU is Incorrect The procedure listed in the service documentation to adjust the TCU from Remote to Local mode is incorrect for Signa Ovation 1 through 4. Both service CD's, part numbers 2283085 Rev.7 and 5112585 Rev.2, (and prior revs.) contain the same incorrect information.

Applies to

Models
Ovation, Signa
Product
Ovation with EXCITE
Software
OPENMR4-MFO4 Applies to Ovation4 (Excite) only. Ovation4 (MFO4) is 11.x based software.
Area
MR

Full cause & step-by-step resolution