GE Healthcare · CTLightSpeedVCT260134637260134602260125037
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Cradle unlocked indicator flashes and wont let the table move See Long Discription for Details Cradle unlocked indicator flashes and wont let the table move
Error message
Error: 260134637 Cradle cannot move because cradle drive is not available or has errors See long descripton Sun Jul 4 08:31:25 2010 Error: 260134637 Host: tgp Process: tgp_subsys File: gtcb_cmc_profile.cxx Line: 95 Sun Jul 4 08:31:24 2010 Host : Table Sub-System Ermes # : 260134637 Exception Class : Secondary Severity : Pri/Soft File : gtcb_cmc_profile.cxx Line# : 95 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 . Cradle cannot move because cradle drive is not available or has errors. Please check cable or drive. Error: 260134602 Cradle drive torque setting error See Long Description for error example Mon Dec 26 10:41:58 2005 Error: 260134602 Host: tgp Process: tgp_subsys File: gtcb_cmc_drv_req.cxx Line: 928 Mon Dec 26 10:41:58 2005 Host : Table Sub-System Ermes # : 260134602 Exception Class : Secondary Severity : Pri/Soft File : gtcb_cmc_drv_req.cxx Line# : 928 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Cradle drive torque setting error Error: 260125037 Table is in shutdown state, cycling power for the table See the LD for the complete error code Error: 260125037 Table is in shutdown state, cycling power for the tabl
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5306599 (replaces 5132669-2 and 5132669-3) Step Motor Driver Positioning GT P - used on GT and VT tables See Long Description for details Part #: 5306599 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: STEP MOTOR DRIVER Positioning GT FRU: Yes Status: Orderable Comments: Per Kittyhawk RSPL, 212008898, 6/9/08 Part #: 5132669-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: STEP MOTOR DRIVER Positioning GT FRU: Yes Status: Orderable Part #: 5132669-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: STEP MOTOR DRIVER Positioning GT FRU: Yes Status: Orderable Comments: added per to PET Table RSPL. source pole/planner code revised per to Global Table transfer
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.