Zipper Artifacts in Shoulder Images and UTNS Notch Diagnostic Fails with Error Code 2247378 due to Uncovered Holes in Pe…

Zipper Artifacts in Shoulder Images and UTNS Notch Diagnostic Fails with Error Code 2247378 due to Uncovered Holes in Penetration Panel.

Symptom

MRIQ Zipper Artifacts in Shoulder Images and UTNS Notch Diagnostic Fails with Error Code 2247378 See Long Description The noise spike counters in the RRF chassis incremented during UTNS diagnostics when the UTNS was not injecting large test spikes.

Error message

2247378 The noise spike counters in the RRF chassis incremented during UTNS diagnostics when the UTNS was not injecting large test spikes. The UTNS or DIF boards might be notching the RF signal unnecessarily. See Long Description for the complete error message Error: 2247378 Host: signa3 Process: diagCoord File: rrfUtns3.c Line: 406 Diagnostic: UTNS Notch Iteration: 1 of 1 Action: 1 Task ID: 0x7f541f0 Time: 104577 Object: APS Test Task Function: Software/Firmware/Config: N/A The noise spike counters in the RRF chassis incremented during UTNS diagnostics when the UTNS was not injecting large test spikes. The UTNS or DIF boards might be notching the RF signal unnecessarily. 487 spikes detected on RF channel 1, expected 0.-> Extended Error: The UTNS can chop out a noise spike on the receiver channels when it sees a spike larger than some preset level. This UTNS detected spikes when it should not have seen any spikes. The board could be faulty or perhaps something is injecting noise into the system. Verify that the room door is closed or disconnect the UTNS antennas from the UTNS boards while running this test.

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.