GE Healthcare · CTLightSpeed260118283260118223500742600074601281683845
LightSpeed 5.X (80/100kW)
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260118283 (25-0593) LVPS 400: Power Suppy output set to off. see long description Tue Jul 1 17:20:59 2008 Error: 260118283 Host: orp Process: orp_subsys File: gen_exec.c Line: 2414 Tue Jul 1 17:20:58 2008 Host : Jedi Ermes # : 260118283 Exception Class : Secondary Severity : Pri/Soft File : gen_exec.c Line# : 2414 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 (25-0593)LVPS 400: Power Suppy output set to off. 3 over/under voltage FW errors in the same LVPS power supply output. Ext15V=0, Ext160V=1, Can15V=2, Gate24V=3, Main_inv=4, Heater1=5, Heater2=6, Error: 260118223 (60-0402)mA scale error See Long Description for complete error Error 260118223 orp orp_subsys gen_exec.c 2213 Host : Jedi Ermes # : 260118223 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2213 Function : No System Function Reported Scan Type : Static Scan: 50074/1/9 (60-0402) mA scale error. mA has been measured to be either too low or too high with respect to mA demand 20 ms after the beginning of the exposure. Z6,Z7= 0h,0h Error: 260007460 Scan ABORT -- Beam Tracking Detected Signal Corruption.^B See LD for complete error Error: 260007460 Host: DAS Proces
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2324118 JH4 AUXILIARY EQUIPED V2 - Used on HP and Brightspeed gantries. Not used on VCT See long description for details Part #: 2324118 Part Type: GP CONSUMABLES Part Weight: -98999.01 LBR Repairable: No Returnable: No Hazardous Material: N Description: JH4 AUXILIARY EQUIPED V2 FRU: Yes Status: Orderable Comments: FLS - JH4 AUXILIARY EQUIPED V2. TO BE USED PER HPOWER PROGRAM RSPL. MS20544 ON 2/12/03.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required