Pular para o conteúdo principal
GE Healthcare · CTLightSpeed260007840

LightSpeed 5.X (60kW)

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260007840 DCB-to-DHCB Communications Failed See long description for full error Tue Nov 16 11:59:22 2004 Error: 260007840 Host: DAS Process: DCB Firmware File: dhcbcomm.c Line: 459 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 4801/2/0 Type: None/Unknown Exception Level: Pri/Soft Time: 11:59:22:975 Log Series: 595 DCB-to-DHCB Communications Failed - Exhausted Comm retries. The DCB Sent The following Command to the DHCB. DHCB Temperature Information Query Command. The Failures are as follows: DHCB UART sem4: 0 DHCB Packet ready sem4: 3 DHCB Nack response: 0 DHCB Packet size mis-match: 0

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2361853-2 GDAS DCB 16-SLICE replaces 2361853 See LD for more Info 2361853-2 GDAS DCB 16-SLICE replaces 2361853 Part #: 2361853-2 Part Type: GP CONSUMABLES Part Weight: -59999.40 LBR Repairable: No Returnable: No Hazardous Material: N Description: GDAS DCB FRU: Yes Status: Orderable FLS - GDAS DCB (16-SLICE) USED ON: LIGHTSPEED PRO 16 GDAS ADDED PER GDAS RSPL MS13452 This board has been replaced by 2361853-2, FE NOTE: Site MUST have M4FMI completed before this board will function properly, ms13452
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required