GE Healthcare · MRSigna22253822225253
Getting SPI-CPB Errors While Prescanning.
Symptom
Getting SPI-CPB Errors While Prescanning. Recieved SPI - CPB errors while prescaning. Looked like the system was not pulsing. Tried software reset, problem remained. The CPB stands for Console Processor Board.
Error message
2225382 SPI-CPB Communication Error. Error: 2225382 Host: oc001 Process: NSP File: HPC_PORT_ID:spi_cpb Line: 300 SPI-CPB Communication Error. The SPI link to the CPB failed - failure detected via timeout. DUART status: 0x0C Last value read from vector/level register: 0x08 Interrupt mask register: 0xDD 2225253 SPI-CPB link is not working, see log.
Applies to
- Models
- Signa
- Product
- Profile 3 Ultrasparc CPU with AN8290Y gradients
- Software
- 7.6x Profile1 through 3 and Contour1 through 3
- Component
- 2185062-2 CPU 486 board 2185062-2 is EOPL. This number has crossed to 2366275-2.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.