Pular para o conteúdo principal
GE Healthcare · MRDiscovery222466622238172270058226760122569252267892

Fails TPS Reset when booting up to Applications or when Resetting TPS due to Defective Y-Axis XPS Control Board SFP (fib…

Fails TPS Reset when booting up to Applications or when Resetting TPS due to Defective Y-Axis XPS Control Board SFP (fiber module).

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2224666 Auto Prescan failed, Gradient amplifiers are not ready

2223817 TPS is Not Responding. Please Reset TPS Wed Apr 28 16:58:01 2004 Error: 2223817 Host: GEMS Process: NSP File: TardisReset.cc Line: 398 TPS is not responding. Please reset TPS.

2270058 The VRF detected that MR Signal is too high (overrange). See Long Description Error: 2270058 Host: signa2 Process: LxMxConsumer File: VrfTrace.cpp Line: 89 Server Name: ADS The VRF detected that MR Signal is too high (overrange).

2267601 AGP forced TPS_reset to fail because AGP could not successfully calibrate the Exciters, Receivers, and Rf-Hub. Look for calibration related error messages in this log See Long Description 2267601 ERROR: EM_IRF3_CAL_FAILED AGP forced TPS_reset to fail because AGP could not successfully calibrate the Exciters, Receivers, and Rf-Hub. Look for calibration related error messages in this log. Extended Error : TPS_reset will fail if the AGP can not calibrate the TG tables for the exciters, the DC offset of the receivers, and the R1 gain of the RF Hub.

2256925 StartupCondition error on SCP resulted in TPS reset failure. See Long Description 2256925 MOI_MR LxMxConsumer services/initialization/ServerInit

Applies to

Models
Discovery
Product
Discovery MR450 Also called DVMR. This product is compatible with 1.5T 20.1, 20.1_IB1, IB2, 21.0 and DV22.0 software systems only.
Software
20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
Component
5183329-2 Fiber Optic Transceiver SFP 4Gb Multimode, RoHS-Compliant Note, as of March 24, 2010 this part can not be ordered yet. It will be a FRU. If you cant order the transceiver order the higher FRU.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues