Fails to Boot Applications Software and TPS reset Fails Due to Corrupt Software.
Symptom
Unable to boot Applications software due to a Host monitor Timeout after 250 seconds See Long Description This was a running system. The tech decided to reboot the system, and then it would never come all the way back up again. Applications software would not come up. After a while, the message "Host monitor Timeout after 250 seconds" was displayed. By viewing the message log from a C-shell using the commands shown below, we found the errors shown below. These same two errors were logged on every failed applications software boot attempt. The two messages are reporting the process "ifcc". Knowing that "ifcc" stands for "Image File Creation and Control", we suspected that the image database was corrupt and would need recovery. We suspected the system was unable to mount the image database completely, and this could be the reason that Applications software would not boot. Looking at the disk partitions using the command "df -k", the /export/home1 image partition was seen to be present (it was 56% full) but perhaps it was corrupt. cd /usr/g/service/log [enter] more gesys* [enter] Fri Sep 1 11:31:19 2006 200002372 signa3 hostMonitor ssProcStart.c 927 The process ifcc took too long to start up. Fri Sep 1 11:32:04 2006 200002352 signa3 hostMonitor ssProcStop.c 812 A HOST_SHUTDOWN was initiated by the death of ifcc.
Error message
200002372 The process ifcc took too long to start up. Fri Jun 10 08:09:19 2005 Error: 200002372 Host: lx-mr Process: hostMonitor File: ssProcStart.c Line: 904 The process ifcc took too long to start up.
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.