GE Healthcare · CTDiscoveryVCT260134715834158
Table will not move down; tape switch was either contacted or not present.
Symptom
Table will not move down; with tape or touch switch error See Long Description for switch locations. Table will not move down; tape switch was either contacted or not present The VCT table has a large number of touch sensors that can produce a variety of error codes.
Error message
Error: 260134715 Elevation touch switch3 (switch on underside of top rear cover) is contacted. See Long Description for complete error Mon Nov 21 14:14:07 2005 Error: 260134715 Host: tgp Process: tgp_subsys File: gtcb_emc_hardware.cxx Line: 714 Mon Nov 21 14:14:07 2005 Host : Table Sub-System Ermes # : 260134715 Exception Class : Secondary Severity : Pri/Soft File : gtcb_emc_hardware.cxx Line# : 714 Function : No System Function Reported Scan Type : None/Unknown Scan: 367/2/0 Elevation touch switch3 (switch on underside of top rear cover) is contacted. Position is 834158 um from the table base.
Applies to
- Models
- Discovery, VCT
- Product
- Discovery CT750 HD V40_PS_HD64_G_GTL V40_PS_HD64_G_GTM Discovery HD CT750 Method of visual identification HD gantry: Gray top and Gray Side covers with HD Markings Console Type: GOC6 has black SCIM and tower face Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of softwares the product can run (Consult attached FCT Software Cross Reference): 08MW39.4 - M3 (see attachement for M3 available options) 09MW29.14 - M4 (HD Options = M3 plus Gemstone Spectral Imaging, Gemstone Spectral Review and volume Helical Shuttle ) Example: Of difference with options software [insite@ct01 ~]$ swokinstall -p Product Software Option Keys: Gemstone Spectral Imaging Permanent HelicalShuttle Permanent Xtream Recon Permanent FAQ's Table weight limit: 500 lbs Tube Type - D3702T Green D3701T Current production - M4 Refer to product data sheets or Technical reference Manual, English 5317222-1EN Revision : 2 or later Refer to Application Tips and Work-Arounds Direction 5311815-1EN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.