GE Healthcare · MRSigna2225382
Mouse Not Working Due To WIM
Symptom
Mouse and Keyboard Quit Working See Long Description
Error message
2225382 - SPI-Workspace Interface Module Communication Error Error: 2225382 Host: SMRITR1 Process: NSP File: HPC_PORT_ID:spi_cpb_wdog_fail Line: 3 SPI-Workspace Interface Module Communication Error. The SPI link to the Workspace Interface Module failed - failure detected via timeout. DUART status: 0x0C Please Reset TPS. If error is persistent, disconnect and reconnect the power cable to the WIM. If error still exists, replace the WIM.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 2141978-6 WORKSPACE INTERFACE MODULE (WIM)
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.