Scanner Hardware Reset Failed.
Symptom
Scanner Hardware Reset Failed. See LD Scanner Hardware Reset Failed Hardware Failed to reset message appeared on monitor AFTER boot-up WITH Error:178568 Failed to establish socket connection to STC in 10 seconds and errors 244553 and 244554 and power up self test fails A, B, and 9 on STC CPU board The following errors were in the gesyslog Thu Jan 13 04:18:32 2005 Error: 178568 Suite: RP25 Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to STC in 10 seconds Thu Jan 13 04:18:32 2005 Error: 244553 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 429 The STC controller has failed to reset. Thu Jan 13 04:18:32 2005 Error: 244554 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 442 The reset failure of the OBC may be due to the reset failure of the STC. Also saw failures A, B, and 9 on the power-up test on the STC CPU .
Error message
Error: 260116316 Communications are DOWN between subsystems See LD for full error Mon Jun 13 14:34:09 2005 Error: 260116316 Host: tgp Process: tgp_subsys File: lan_watchdog.c Line: 904 Mon Jun 13 14:34:03 2005 Host : Table Gantry Processor Ermes # : 260116316 Exception Class : Safe State Severity : Primary File : lan_watchdog.c Line# : 904 Function : No System Function Reported Scan Type : None/Unknown Scan: 7754/5/0 Communications are DOWN between subsystems Or Sun Dec 17 08:26:45 2006 Error: 260116316 Host: tgp Process: tgp_subsys File: lan_watchdog.c Line: 996 Sun Dec 17 08:26:39 2006 Host : Table Gantry Processor Ermes # : 260116316 Exception Class : Safe State Severity : Secondary File : lan_watchdog.c Line# : 996 Function : No System Function Reported Scan Type : None/Unknown Scan: 50119/1/0 Slip-Ring Communications are DOWN between subsystems Error: 260116307 Could not send message to 101 or 201 from 101 or 201 (any combination) see long description for error details. Error: 260116307 Could not send message to 101 or 201 from 101 or 201. See examples below: Fri Aug 15 15:01:01 2008 Error: 260116307 Host: tgp Process: tgp_subsys File: lan_watchdog.c Line: 513 Fri Aug 15 15:0
Applies to
- Models
- LightSpeed
- Product
- LightSpeed RT4 WB O G4 G HPT WB O M4 G HPT LightSpeed RT4 with GRE and MDAS4 or GDAS4 gantry At GE Medical Systems, we’re pioneering exciting CT technology for Radiation Therapy planning. Introducing the newest addition to our LightSpeed® series – LightSpeed RT, the first multi-slice, wide-bore CT scanner with the largest field of view in the industry. With a generous 80 cm gantry opening, the LightSpeed RT bore is 10 cm larger than conventional scanners. This provides complete flexibility for patient positioning freedom – even for the most demanding oncology needs. And the 65 cm variable field of view delivers 15 additional centimeters of visual coverage, allowing clinicians to see a patient’s entire anatomy for precise radiation therapy planning. Position and precision Thin, multi-slice imaging is the extra bonus – enabling respiratory gating, providing precise target delineation and delivering higher resolution digitally reconstructed radiographs (DRRs) than ever before. All adding up to the ability for you to deliver what really matters most – better cancer care for your patients. Plus, the LightSpeed RT scanner is available with the Exact™ couch insert, enabling you to consistently replicate patient immobilization for treatment. With LightSpeed RT, the future of your Radiation Oncology department is wide open. Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2389885-3 (replaces 2389885-2 and 2290254-3) ORP2 BOARD ASSY Positioning HP60 See Long Description for details Part #: 2389885-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: ORP2 BOARD ASSY Positioning HP60 FRU: Not Available Status: Orderable Comments: Per ECR 2036188, 212008898, 12/6/2006. Source Pole/Planner Code revised per to supplier confirmation by sso#305010201 2389885-2 ORP2 - Used on LightSpeed 5.X and BrightSpeed (Replaces 2290254-3 per PQR 13030308 ) Part #: 2389885-2 Part Type: GP CONSUMABLES Part Weight: 2.20 LBR Repairable: No Returnable: No Hazardous Material: N Description: ORP2 BOARD ASSY Positioning HP60 FRU: Yes Status: Orderable Comments: Not Available Part #: 2290254-3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: P2033WC ORP BD ASSY FRU: Yes Status: Orderable FLS - P2033WC ORP BD ASSY USED ON H-POWER, ORP ASSY
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.