GE Healthcare · X-ray R&F
error 3412, MPH 3412 error, Prestige, GPX
Symptom
error 3412, MPH 3412 error, Prestige, GPX getting error 3412, found MPH 5vdc supply low, was 4.76vdc. Adjust 5vdc supply for 5.06.
Applies to
- Product
- MPH
- Component
- 2157127:cable W27(from A5PS1 LV1 power supply to Command 1 board) replace the cable W27
- Area
- X-ray R&F
Full cause & step-by-step resolution