error 3412, MPH 3412 error, , GPX

Symptom

error 3412, MPH 3412 error, Prestige, GPX getting error 3412, found MPH 5vdc supply low, was 4.76vdc. Adjust 5vdc supply for 5.06.

M3412,replace the cable from A5PS1 LV1 POWER SUPPLY check the oput of A5PS1 LV1 POWER SUPPLY is ok,but the DC5V on PU-CPU is low.replace the cable from A5PS1 LVA POWER SUPPLY to Command 1 board.CABLE W27:2157127

Applies to

Product
Prestige SI
Area
X-ray R&F

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.